Measurement of carboxyhemoglobin and total hemoglobin by five specialized spectrophotometers (CO-oximeters) in comparison with reference methods

JJ Mahoney, HJ Vreman, DK Stevenson… - Clinical …, 1993 - academic.oup.com
JJ Mahoney, HJ Vreman, DK Stevenson, AL Van Kessel
Clinical Chemistry, 1993academic.oup.com
We measured total hemoglobin (CtHb) and carboxyhemoglobin (COHb) in 100 patients'
blood samples by using five specialized spectrophotometers (CO-oximeters)--IL 482 CO-
Oximeter, Corning 2500 CO-oximeter, Radiometer OSM 3 Hemoximeter, Corning 270 CO-
oximeter, and the AVL 912 CO-Oxylite--and compared the results with those obtained with
the manual cyanmethemoglobin method and a gas-chromatographic (GC) method,
respectively. For the CtHb measurements, the differences between the cyanmethemoglobin …
Abstract
We measured total hemoglobin (CtHb) and carboxyhemoglobin (COHb) in 100 patients' blood samples by using five specialized spectrophotometers (CO-oximeters)--IL 482 CO-Oximeter, Corning 2500 CO-oximeter, Radiometer OSM 3 Hemoximeter, Corning 270 CO-oximeter, and the AVL 912 CO-Oxylite--and compared the results with those obtained with the manual cyanmethemoglobin method and a gas-chromatographic (GC) method, respectively. For the CtHb measurements, the differences between the cyanmethemoglobin method and the CO-oximeters were not clinically important for any model. For the blood COHb measurements, the direction of the bias relative to GC was dependent on COHb concentration. In general, the CO-oximeters underestimated COHb concentration for COHb > 2.5% of total hemoglobin but overestimated COHb concentration for COHb < or = 2.5%. We conclude that all five CO-oximeters compared favorably with the reference methods for CtHb and for high concentrations of COHb. However, the inaccuracy of CO-oximeters for low-concentration (< or = 2.5%) COHb measurements may make these instruments unsuitable for some applications.
Oxford University Press