A new PET scanner with semiconductor detectors enables better identification of intratumoral inhomogeneity

T Shiga, Y Morimoto, N Kubo, N Katoh… - Journal of Nuclear …, 2009 - Soc Nuclear Med
T Shiga, Y Morimoto, N Kubo, N Katoh, C Katoh, W Takeuchi, R Usui, K Hirata, S Kojima…
Journal of Nuclear Medicine, 2009Soc Nuclear Med
An autoradiography method revealed intratumoral inhomogeneity in various solid tumors. It
is becoming increasingly important to estimate intratumoral inhomogeneity. However, with
low spatial resolution and high scatter noise, it is difficult to detect intratumoral
inhomogeneity in clinical settings. We developed a new PET system with CdTe
semiconductor detectors to provide images with high spatial resolution and low scatter
noise. Both phantom images and patients' images were analyzed to evaluate intratumoral …
An autoradiography method revealed intratumoral inhomogeneity in various solid tumors. It is becoming increasingly important to estimate intratumoral inhomogeneity. However, with low spatial resolution and high scatter noise, it is difficult to detect intratumoral inhomogeneity in clinical settings. We developed a new PET system with CdTe semiconductor detectors to provide images with high spatial resolution and low scatter noise. Both phantom images and patients' images were analyzed to evaluate intratumoral inhomogeneity.
Methods
This study was performed with a cold spot phantom that had 6-mm-diameter cold sphenoid defects, a dual-cylinder phantom with an adjusted concentration of 1:2, and an “H”-shaped hot phantom. These were surrounded with water. Phantom images and 18F-FDG PET images of patients with nasopharyngeal cancer were compared with conventional bismuth germanate PET images. Profile curves for the phantoms were measured as peak-to-valley ratios to define contrast. Intratumoral inhomogeneity and tumor edge sharpness were evaluated on the images of the patients.
Results
The contrast obtained with the semiconductor PET scanner (1.53) was 28% higher than that obtained with the conventional scanner (1.20) for the 6-mm-diameter cold sphenoid phantom. The contrast obtained with the semiconductor PET scanner (1.43) was 27% higher than that obtained with the conventional scanner (1.13) for the dual-cylinder phantom. Similarly, the 2-mm cold region between 1-mm hot rods was identified only by the new PET scanner and not by the conventional scanner. The new PET scanner identified intratumoral inhomogeneity in more detail than the conventional scanner in 6 of 10 patients. The tumor edge was sharper on the images obtained with the new PET scanner than on those obtained with the conventional scanner.
Conclusion
These phantom and clinical studies suggested that this new PET scanner has the potential for better identification of intratumoral inhomogeneity, probably because of its high spatial resolution and low scatter noise.